Pendulum Impact Testing Machine 50J Auto rasie and brake Charpy impact test ISO 179 PIT501J-TS
Functions
PIT series j-type plastic pendulum impact testing machine is mainly for testing the impact resistance of non-metallic materials such as plastics, nylon, rubber, glass fiber reinforced plastics, composite plastic pipes and electrical insulation materials under dynamic load.
The machine can be divided into touch screen type and microcomputer type according to the display mode. It is easy to operate and has high working efficiency.
It can do Charpy, Izod and tensile impact test. Support adjustment and replacement is convenient, strong applicability. Equipped with tensile impact pendulum and fixtures, it can carry on tests on plastic film and sheet.
It is the ideal instruments for hard plastic manufacturers, plastic pipe manufacturers, quality inspection units, scientific research institute.
Standards
ISO 179, ISO 180, ISO 8256, ISO 9854.1, ASTM D256, ASTM D1822, ASTM D6110
Features
High resolution
Using a shaftless encoder allows angle resolution of 0.045 °
Automatic hammer uplift
Automatic hammer rising to pre-set angle after finish test.
Settable impact angle
Angle range 30°~150°, automatic calculate out the result of absorbed energy according setting angle.
Auto-brake system
Automatically brake after impact test and power outage, safe and high efficient.
Energy losses
Shaftless encoder for angle measurement rotates without any friction; only support bearings friction and windage has the lowest energy losses
Touch screen
Touch screen provides easy setup of specimen dimensions, pendulum energy, units and calibration data. Also it can intuitionally display test results: absorbed energy and impact toughness
Exchangeable pendulum
Simply change the pendulum to satisfy Charpy, Izod and tensile impact test
Center alignment
Rotating design is quick and accurate to align specimen center for Charpy test
Supports and anvils
Supports are tightened to seat by dovetail groove firmly; only need remove two bolts to change the supports
Parameter
1. Capacity
3. Angle resolution: 0.045°
- Charpy Test for plastics
Items | ISO 179, GB/T 1043 | ASTM D6110 |
Angle of striking | 150° | 150° |
Velocity of striking | 3.8m/s | 3.46m/s |
Distance from the axis of support to the center of percussion | 395mm | 327mm |
Span | 40mm, 60mm, 70mm (can be adjusted by span block) | 95.3mm |
Radius of striker edge | 2±0.5mm | 3.17±0.12mm |
Specimen dimension (LxWxH) | 80x10x4mm | 127x12.7x(3~12.7)mm |
- Charpy Test for plastic pipe
Standard | ISO 9854.1 |
Angle of striking | 150° |
Velocity of striking | 3.8m /s |
Distance from the axis of support to the center of percussion | 395mm |
Span | 40mm, 60mm, 62mm, 70mm |
Radius of striker edge | 3.17±0.12mm |
Specimen dimension Length x width x thickness (mm) | Type | Length | Width | thickness | Support span |
1 | 100±2 | Whole pipe | 70±0.5 | ||
2 | 50±1 | 6±0.2 | e | 40±0.5 | |
3 | 120±2 | 15±0.5 | e | 70±0.5 | |
Note: e is the machining thickness of pipe |
- Izod test
Standard | ISO 180 | ASTM D256 |
Angle of striking | 150° | 145.2° |
Velocity of striking | 3.5m/s | 3.46m/s |
Distance from the axis of support to the center of percussion | 335 | |
Striking knife center to vice upper surface | 22±0.2mm | |
Radius of striker edge | 0.8±0.2mm | |
Specimen dimension (LxWxH) | 80×10×4mm, 63.5×12.7×12.7mm 63.5×12.7×6.4mm, 3.5×12.7×3.3mm |
- Tesile impact test
Standard | ASTM D1822 | ISO 8256 |
Angle of striking | 150° | 150° |
Velocity of striking | 3.46m/s | 3.8m/s |
Distance from the axis of support to the center of percussion | 327mm | 395mm |
Crosshead mass | 60±1g, 120±1g | |
Maximum specimen thickness | 4mm |
8. Power supply: 220V±10% 50Hz 500W
9. Machine dimension (LxWxH): 1000mm×300mm×760mm
10. Weight: 300kg